lamp_export.py writes the per-lamp bundle consumed by the buildfor.life comparison pages: spd.csv (full 380-1050 nm spectrum), tm30.csv (ANSI/IES TM-30-18 hue-bin data for the color vector graphic), and metrics.json (photometric, colorimetric, CRI R1-R15, TM-30 Rf/Rg, electrical, supply settings). - tm30.py computes TM-30-18 via colour-science from the measured spectrum, resampled to a uniform 1 nm grid; validated against CIE FL2 (Rf 70, Rg 86) - Optional built-in PSU control (--mode/--voltage/--frequency/--current), switched on for the measurement and off afterwards, with --settle warm-up and --interval between readings - Readings use the vendor single-shot cycle verified from USB captures: test config -> trigger (8C 0E 02) -> poll -> read -> reset per reading - Also supports --passive and --parse <pcap> sources
HPCS 6500 — Open-Source Driver
Python driver and CLI for the HPCS 6500 spectrophotometer / integrating sphere.
Communicates directly with the instrument over USB serial, replacing the vendor software for measurement automation and data extraction.
What This Does
- Measure: Luminous flux (lm), CCT (K), CRI (Ra, R1-R15), chromaticity (CIE xy, uv, u'v'), radiometric flux, spectrum 380-1050 nm
- Electrical: Voltage, current, power, power factor, frequency
- Harmonics: 50-harmonic voltage/current analysis, UThd, AThd, waveforms
- Power supply: Control the built-in AC (100-240V, 50/60Hz) and DC (1-60V, 0-5A) power supply
- Export: CSV output for data logging
- TM-30: ANSI/IES TM-30-18 Rf, Rg, and hue-bin data computed from the measured spectrum (via colour-science)
- Lamp bundles: one-command export of spectrum + TM-30 + metrics for the buildfor.life lamp comparison
Quick Start
# Install
uv sync
# Single measurement (auto-detects device, turns PSU on/off)
uv run hpcs6500.py
# Continuous measurements
uv run hpcs6500.py --continuous
# Quick test (lumen + CCT only)
uv run hpcs6500.py --quick
uv run hpcs6500.py --quick --continuous
# Show full spectrum + harmonics
uv run hpcs6500.py --spectrum --harmonics
# Save to CSV
uv run hpcs6500.py --continuous --csv output.csv
Power Supply Control
# Read current PSU settings
uv run hpcs6500.py --psu-status
# Set AC mode, 230V 50Hz
uv run hpcs6500.py --mode ac --voltage 230 --frequency 50
# Set DC mode, 12V with 1A current limit
uv run hpcs6500.py --mode dc --voltage 12 --current 1.0
# Manual PSU on/off
uv run hpcs6500.py --psu-on
uv run hpcs6500.py --psu-off
# Set integration time (ms)
uv run hpcs6500.py --integration 500
Lamp Comparison Export
Produces the per-lamp data bundle consumed by the buildfor.life comparison
pages: spd.csv (full 380-1050 nm spectrum), tm30.csv (TM-30-18 hue-bin
data for the color vector graphic), and metrics.json (photometric,
colorimetric, CRI R1-R15, TM-30 Rf/Rg, electrical).
# Single reading from the device
uv run lamp_export.py --name philips-a60-8w --manufacturer Philips --model "A60 8W 927"
# Average several readings
uv run lamp_export.py --name philips-a60-8w --readings 5
# Power the lamp from the built-in supply: 230 V / 50 Hz, 60 s warm-up,
# PSU switches on before the readings and off afterwards
uv run lamp_export.py --name philips-a60-8w --voltage 230 --frequency 50 --settle 60
# From an existing pcap capture
uv run lamp_export.py --name some-lamp --parse captures/run.pcap
Output lands in lamps/<name>/. TM-30 is computed from the measured spectrum
with colour-science; the spectrum is
relative, which TM-30 is invariant to. Sanity check of the implementation:
uv run tm30.py reproduces the published values for the CIE FL2 illuminant
(Rf 70, Rg 86).
Offline Parsing
Parse previously captured USB traffic (pcap files from USBPcap):
uv run hpcs6500.py --parse captures/some_capture.pcap
uv run hpcs6500.py --parse captures/some_capture.pcap --quick
Files
| File | Description |
|---|---|
hpcs6500.py |
Driver class (HPCS6500) and CLI |
lamp_export.py |
Lamp comparison bundle export (spd/tm30/metrics) |
tm30.py |
ANSI/IES TM-30-18 computation from a spectrum |
usb_capture.py |
USB traffic capture tool (requires USBPcap) |
PROTOCOL.md |
Complete protocol reference (byte-level) |
pyproject.toml |
Project metadata and dependencies |
Hardware
- Device: HPCS 6500 spectrophotometer / integrating sphere
- USB: STM32 Virtual COM Port (VID
0483, PID5741) - Protocol: Custom binary over serial, documented in PROTOCOL.md
Dependencies
- Python 3.11+
pyserial(serial communication)colour-science(TM-30 computation)- USBPcap (only for
usb_capture.py, not needed for normal operation)
Protocol
The binary protocol is fully documented in PROTOCOL.md, including all command bytes, data block layouts, field offsets, and the complete measurement sequence. This was reverse-engineered from USB packet captures of the vendor software.