# HIOKI 3193-10 Power Analyzer Tools
|
|
|
|
GPIB control tools for the HIOKI 3193-10 power analyzer, built for solar MPPT converter efficiency testing.
|
|
|
|
## Hardware Setup
|
|
|
|
- **HIOKI 3193-10** power analyzer with 9600 input units
|
|
- **GPIB-USB adapter**: [UsbGpib](https://github.com/xyphro/UsbGpib) — open-source USB-GPIB adapter, compatible with pyvisa (appears as a USB-TMC device)
|
|
- **NI-VISA runtime** — provides the VISA layer (`visa32.dll`)
|
|
|
|
The UsbGpib adapter connects to the HIOKI's rear GPIB port and is recognized by NI-VISA as a standard USB-TMC instrument, so it works seamlessly with pyvisa.
|
|
|
|
## Requirements
|
|
|
|
- Python 3.12+
|
|
- [uv](https://docs.astral.sh/uv/) package manager
|
|
- NI-VISA runtime (already installed if `visa32.dll` exists)
|
|
- UsbGpib adapter connected between PC and HIOKI 3193-10
|
|
|
|
## Install
|
|
|
|
```bash
|
|
uv sync
|
|
```
|
|
|
|
## Quick Start
|
|
|
|
```bash
|
|
# 1. Verify connection
|
|
uv run hioki identify
|
|
|
|
# 2. Configure for MPPT testing (Ch5=solar input, Ch6=MPPT output)
|
|
uv run hioki setup-mppt
|
|
|
|
# 3. Take a single reading
|
|
uv run hioki measure
|
|
|
|
# 4. Monitor continuously (prints to terminal)
|
|
uv run hioki monitor
|
|
|
|
# 5. Monitor with live graph
|
|
uv run hioki live
|
|
```
|
|
|
|
## Commands
|
|
|
|
### `hioki identify`
|
|
|
|
Show instrument identity, installed options, clock, wiring mode, and response speed.
|
|
|
|
### `hioki setup-mppt`
|
|
|
|
One-command setup for MPPT efficiency testing:
|
|
- Ch5: V-auto, I-auto, DC coupling, SLOW (solar panel input)
|
|
- Ch6: V-auto, I-auto, DC coupling, SLOW (MPPT converter output)
|
|
- EFF1 = P6/P5 x 100%
|
|
- Display: Ch5 items on left, Ch6 items on right, with EFF1
|
|
|
|
Both voltage and current auto-range for best accuracy across the full sweep range.
|
|
|
|
### `hioki measure [items...]`
|
|
|
|
Take a single measurement. Default items: `U5 I5 P5 U6 I6 P6 EFF1`
|
|
|
|
```bash
|
|
# Default (MPPT channels)
|
|
uv run hioki measure
|
|
|
|
# Custom items
|
|
uv run hioki measure U1 I1 P1 FA
|
|
|
|
# All 6 channels
|
|
uv run hioki measure U1 I1 P1 U2 I2 P2 U3 I3 P3 U4 I4 P4 U5 I5 P5 U6 I6 P6
|
|
```
|
|
|
|
### `hioki monitor [items...]`
|
|
|
|
Continuous text-based monitoring with optional CSV logging. When logging to CSV, voltage and current ranges are recorded per sample.
|
|
|
|
```bash
|
|
# Default 1-second interval
|
|
uv run hioki monitor
|
|
|
|
# Fast polling, 100 readings, save to CSV
|
|
uv run hioki monitor -i 0.5 -n 100 -o data.csv
|
|
|
|
# Monitor specific items
|
|
uv run hioki monitor P5 P6 EFF1
|
|
```
|
|
|
|
Options:
|
|
- `-i, --interval` - Seconds between readings (default: 1.0)
|
|
- `-n, --count` - Number of readings, 0 = infinite (default: 0)
|
|
- `-o, --output` - CSV output file path (includes U/I range columns)
|
|
|
|
### `hioki live [items...]`
|
|
|
|
Real-time graphing monitor. Opens a matplotlib window with live-updating plots grouped by type (voltage, current, power, efficiency).
|
|
|
|
```bash
|
|
# Default view
|
|
uv run hioki live
|
|
|
|
# Power and efficiency only, log to file
|
|
uv run hioki live -o test_log.csv P5 P6 EFF1
|
|
|
|
# Fast updates, shorter history
|
|
uv run hioki live -i 0.5 --history 100 U5 U6 P5 P6 EFF1
|
|
```
|
|
|
|
Options:
|
|
- `-i, --interval` - Seconds between readings (default: 1.0)
|
|
- `-o, --output` - CSV output file path (includes U/I range columns)
|
|
- `--history` - Max data points on graph (default: 300)
|
|
|
|
### `hioki display-select [items...]`
|
|
|
|
Configure the instrument's front panel display.
|
|
|
|
```bash
|
|
# MPPT preset: Ch5 left, Ch6 right, EFF1
|
|
uv run hioki display-select --mppt
|
|
|
|
# Custom 16-item layout
|
|
uv run hioki display-select U5 I5 P5 U6 I6 P6 EFF1
|
|
```
|
|
|
|
Options:
|
|
- `-c, --count` - Number of display slots: 4, 8, or 16 (default: 16)
|
|
- `--mppt` - Preset: Ch5 on left, Ch6 on right, with EFF1
|
|
|
|
### `hioki efficiency`
|
|
|
|
Configure efficiency calculation formula.
|
|
|
|
```bash
|
|
# EFF1 = P6/P5 (default from setup-mppt)
|
|
uv run hioki efficiency --numerator P6 --denominator P5
|
|
|
|
# EFF2 for a different pair
|
|
uv run hioki efficiency -f 2 -n P4 -d P3
|
|
```
|
|
|
|
### `hioki integration <action>`
|
|
|
|
Control energy integration (Wh, Ah totals).
|
|
|
|
```bash
|
|
uv run hioki integration start # Reset and start integrating
|
|
uv run hioki integration status # Check which channels are running
|
|
uv run hioki integration stop # Stop and print Wh/Ah/time results
|
|
uv run hioki integration reset # Reset counters
|
|
```
|
|
|
|
### `hioki send <command>`
|
|
|
|
Send raw SCPI/GPIB commands. Queries (containing `?`) return the response.
|
|
|
|
```bash
|
|
uv run hioki send *IDN?
|
|
uv run hioki send :VOLTage5:RANGe?
|
|
uv run hioki send :VOLTage5:RANGe 300
|
|
uv run hioki send *RST
|
|
```
|
|
|
|
## Global Options
|
|
|
|
- `-a, --address` - VISA address (auto-detects HIOKI if omitted)
|
|
- `--timeout` - Communication timeout in ms (default: 5000)
|
|
|
|
## CSV Output Format
|
|
|
|
When using `-o` with `monitor` or `live`, the CSV includes:
|
|
- `timestamp` - Date and time of each reading
|
|
- Measurement values (e.g., `U5`, `I5`, `P5`, `EFF1`)
|
|
- `U<ch>_range` - Active voltage range per channel (e.g., 150, 300)
|
|
- `I<ch>_range` - Active current range per channel (e.g., 0.2, 5, 50)
|
|
|
|
Range columns help identify auto-range transitions and assess measurement accuracy.
|
|
|
|
## Python API
|
|
|
|
Use the driver directly in your own scripts:
|
|
|
|
```python
|
|
from hioki3193 import Hioki3193
|
|
|
|
with Hioki3193("GPIB0::1::INSTR") as meter:
|
|
print(meter.idn())
|
|
|
|
# Configure
|
|
meter.set_wiring_mode("1P2W")
|
|
meter.set_coupling(5, "DC")
|
|
meter.set_voltage_auto(5, True)
|
|
meter.set_current_auto(5, True)
|
|
meter.set_coupling(6, "DC")
|
|
meter.set_voltage_auto(6, True)
|
|
meter.set_current_auto(6, True)
|
|
meter.set_response_speed("SLOW")
|
|
meter.set_efficiency(1, "P6", "P5")
|
|
|
|
# Read
|
|
result = meter.measure("U5", "I5", "P5", "U6", "I6", "P6", "EFF1")
|
|
print(f"Input: {result.values['P5']:.2f} W")
|
|
print(f"Output: {result.values['P6']:.2f} W")
|
|
print(f"Efficiency: {result.values['EFF1']:.1f}%")
|
|
```
|
|
|
|
## Measurement Items
|
|
|
|
| Code | Description | Code | Description |
|
|
|------|-------------|------|-------------|
|
|
| U1-U6 | Voltage per channel | I1-I6 | Current per channel |
|
|
| P1-P6 | Active power | S1-S6 | Apparent power |
|
|
| Q1-Q6 | Reactive power | PF1-PF6 | Power factor |
|
|
| DEG1-DEG6 | Phase angle | PK1-PK6 | Peak value |
|
|
| FA,FB,FC | Frequency | EFF1-EFF3 | Efficiency |
|
|
| U12,P12... | Sum pairs | U123,P123... | Sum triples |
|
|
| WP1-WP6 | Integrated Wh | IH1-IH6 | Integrated Ah |
|
|
| TIME | Integration elapsed time | | |
|
|
|
|
## Error Values
|
|
|
|
When reading data, these special values indicate errors:
|
|
- `+6666.6E+99` - Display blank (no data)
|
|
- `+7777.7E+99` - Scaling error (e.g., division by zero in efficiency)
|
|
- `+9999.9E+99` - Input over range
|
|
|
|
## Reference
|
|
|
|
See `GPIB_COMMAND_REFERENCE.txt` for the complete command reference extracted from the HIOKI 3193-10 manual.
|