Add CP mode sanity check: validate current limit vs power at min voltage
CP=500W at V=50V needs 10A on the load side — if the supply I_limit is only 20A and the MPPT has conversion losses, this can cause the supply to current-limit and timeout. The check now catches this upfront: CP check: 500W / 50V = 10.0A (limit 20.0A) OK Also raises ValueError if the worst-case current exceeds the limit. Co-Authored-By: Claude Opus 4.6 <noreply@anthropic.com>
This commit is contained in:
@@ -564,11 +564,19 @@ class MPPTTestbench:
|
||||
self,
|
||||
max_voltage: float,
|
||||
current_limit: float,
|
||||
*,
|
||||
min_voltage: float | None = None,
|
||||
load_mode: str = "CC",
|
||||
max_load_setpoint: float = 0.0,
|
||||
) -> None:
|
||||
"""Check that the supply can deliver the requested V and I.
|
||||
|
||||
Queries the supply's voltage range (max V) and verifies the
|
||||
requested parameters are within bounds. Raises ValueError if not.
|
||||
requested parameters are within bounds. For CP mode, also
|
||||
checks that the supply current limit can deliver the requested
|
||||
power at the minimum voltage.
|
||||
|
||||
Raises ValueError if the configuration is invalid.
|
||||
"""
|
||||
supply_max_v = self.supply.get_voltage_range()
|
||||
if max_voltage > supply_max_v:
|
||||
@@ -585,6 +593,21 @@ class MPPTTestbench:
|
||||
f"(range {supply_max_v:.0f}V)"
|
||||
)
|
||||
|
||||
# CP mode: at min voltage the load draws max current
|
||||
if load_mode == "CP" and max_load_setpoint > 0 and min_voltage:
|
||||
worst_case_i = max_load_setpoint / min_voltage
|
||||
if worst_case_i > current_limit:
|
||||
raise ValueError(
|
||||
f"CP mode: {max_load_setpoint:.0f}W at "
|
||||
f"{min_voltage:.0f}V needs {worst_case_i:.1f}A "
|
||||
f"but supply I_limit is {current_limit:.1f}A"
|
||||
)
|
||||
print(
|
||||
f" CP check: {max_load_setpoint:.0f}W / "
|
||||
f"{min_voltage:.0f}V = {worst_case_i:.1f}A "
|
||||
f"(limit {current_limit:.1f}A) OK"
|
||||
)
|
||||
|
||||
# ── 2D Voltage × Current Sweep ─────────────────────────────────────
|
||||
|
||||
def sweep_vi(
|
||||
@@ -630,7 +653,14 @@ class MPPTTestbench:
|
||||
|
||||
# Sanity check
|
||||
max_v = max(abs(v_start), abs(v_stop))
|
||||
self.check_supply_capability(max_v, current_limit)
|
||||
min_v = min(abs(v_start), abs(v_stop))
|
||||
max_l = max(abs(l_start), abs(l_stop))
|
||||
self.check_supply_capability(
|
||||
max_v, current_limit,
|
||||
min_voltage=min_v,
|
||||
load_mode=load_mode,
|
||||
max_load_setpoint=max_l,
|
||||
)
|
||||
|
||||
# Auto-correct step directions
|
||||
if v_start < v_stop and v_step < 0:
|
||||
|
||||
@@ -1192,7 +1192,14 @@ class TestbenchGUI(tk.Tk):
|
||||
|
||||
# Sanity check
|
||||
max_v = max(abs(v_start), abs(v_stop))
|
||||
bench.check_supply_capability(max_v, current_limit)
|
||||
min_v = min(abs(v_start), abs(v_stop))
|
||||
max_l = max(abs(l_start), abs(l_stop))
|
||||
bench.check_supply_capability(
|
||||
max_v, current_limit,
|
||||
min_voltage=min_v,
|
||||
load_mode=load_mode,
|
||||
max_load_setpoint=max_l,
|
||||
)
|
||||
|
||||
bench.supply.set_current(current_limit)
|
||||
bench.supply.output_on()
|
||||
|
||||